مشخصه
|
توضیحات
|
حد تشخیص
اسپکتروفتومتر ICP-MS کارکرده
|
Based on three times the standard deviation of the blank using three-second integration time and peak
hopping at 1-point per mass.
Element ng/L (ppt)
9Be < 1
56Fe < 3
59Co < 1
115In < 0.5
238U < 0.5
|
حساسیت اسپکتروفتومتر ICP-MS دست دوم
|
Element M cps/mg/L
9Be > 3
24Mg > 20
115In > 50
238U > 40
|
گونه های اکسید و شارژ مضاعف اسپکتروفتومتر ICP-MS دست دو
|
Measured without the use of a desolvation device such as a chilled spray chamber and under identical
operating conditions used to achieve sensitivity and detection-limit specifications.
CeO+/Ce+ < 0.025
Ce++/Ce+ < 0.03
|
سیگنال زمینه اسپکتروفتومتر ICP-MS استوک
|
Mass 220 < 1 cps
|
دقت کوتاه مدت اسپکتروفتومتر ICP-MS ریفربیش
|
Defined as the relative standard deviation (% RSD) for a 1-10 µg/L multielement solution, automatically cycling
between Standard, DRC and KED modes, using a 3-second integration time, without internal standardization.
< 3% RSD
|
پایداری بلند مدت اسپکترومتر ICP-MS کارکرده
|
Relative stability after a one-hour warm-up period. Defined as the relative standard deviation of the mean signal for a 1-10 µg/L
multielement solution, automatically cycling between Standard, DRC and KED modes, measured once every 10 minutes, without internal standardization.
< 4% RSD over 4 hours
|
دقت نسبت ایزوتوپ اسپکترومتر ICP-MS دست دوم
|
Defined for the isotope ratio of 107Ag/109Ag. Obtained using single-point peak hopping.
< 0.08*% RSD (*or within a factor of two of the counting statistics limit)
|
پایداری کالیبراسیون جرمی اسپکترومتر ICP-MS دست دو
|
Measured using a 1 µg/L multielement solution containing 7Li, 24Mg, 115In and 238U. Defined in terms of the shift in spectral
position corresponding to maximum spectral peak intensity for each element, obtained without the use of multiple-point, peak-searching algorithms.
< 0.05 amu over 8 hours of continuous operation
|
سرعت Quadrupole peak hop (slew)
اسپکترومتر ICP-MS استوک
|
Defined as the maximum rate at which the quadrupole can change from the minimum mass to the maximum mass without affecting the precision of the analytical measurement.
1.6M amu/sec
|
سرعت اسکن Quadrupole اسپکتروسکوپی ICP-MS کارکرده
|
Defined as the maximum rate at which the quadrupole can be scanned while acquiring continuous spectral data at every mass from the minimum to the maximum mass of the instrument (1-285 amu).
5000 amu/sec
|
حساسیت فراوانی اسپکتروسکوپی ICP-MS دست دوم
|
Defined as the intensity of a given isotope at spectral peak maximum, relative to the intensity of that isotope at 1 amu lower and at 1 amu higher than the mass position corresponding to peak maximum.
Measured at 23Na:
Better than 1.0 x 10-6 at low mass side of peak
Better than 1.0 x 10-7 at high mass side of peak
|
محدوده خطی آشکارساز اسپکتروسکوپی ICP-MS دست دو
|
The SimulScan™ detection system operates from < 0.1 cps to > 109 cps. This provides over 10 orders of magnitude of linear dynamic range in a single continuous scan.
|
سرعت دستیابی اطلاعات اسپکتروسکوپی ICP-MS استوک
|
> 3000 temporal data points/sec maximum
|