مشخصه
|
توضیحات
|
آون ستون کروماتوگرافی گازی GC
|
Dimensions: 202 × 200 × 105 mm (HWD).
|
Operating temperature range suitable for all columns and chromatographic separations. Ambient temperature +4 °C to 450 °C.
With LN2 cryo cooling: –80 to 450 °C.
With CO2 cryo cooling: –40 to 450°C
|
Oven power is turned off automatically when the lid is opened
|
Column basket diameter: 130 mm
|
Standard: 5 °C above ambient to 350 °C. With CO2 cryo: –20 °C to 350 °C.
|
Temperature setpoint resolution: 1 °C
|
Column bleed compensation
|
Maximum run time: 999.99 min
|
Temperature programming: six ramps with seven plateaus
|
Ambient rejection: <0.01 °C per 1 °C ambient change
|
Automatic H2 carrier safety turnoff when inlet gas pressure drops
|
Typical cool-down rate: 350 °C to 50 °C: 5.2 min
|
Cryogenic gas consumption is typically half that of dual-channel ovens
|
کنترل الکترونیکی پنوماتیک ها (EPC) کروماتوگراف گازی GC
|
Pressure setpoint increments: 0.01 psi
|
Temperature and pressure sensors compensate for ambient variation and altitude
|
All EPC setpoints are included in the method
|
Maximum number of modules: one
|
Maximum number of channels: three
|
Pressure increments: 0.01 psi
|
Maximum pressure: 100 psig
|
Maximum pressure ramp rate (when configured as Inlet): 150 psi/min
|
ورودی های کروماتوگراف گاز GC
|
Maximum of two inlets installed.
|
Split/Splitless capillary inlet (S/SL)
Purged packed injection port (PPIP)
Temperature–programmable vaporizer (PTV)
Temperature–programmable cool-on-column (PCOC)
Electronic pneumatics control (EPC)
|
When capillary column dimensions are entered, the actual flow and pressure are shown as calibrated digital readouts
|
Carrier gas selection provided for: He, H2, N2 and Ar-CH4
|
Pressure setting range: 0 to 100 psi
|
Constant pressure, constant flow modes
|
Three ramps for pressure or flow
|
Optional Merlin Microseal septum guarantees at least 2,000 injections
|
Septum purge set automatically
|
Split/Splitless گاز کروماتوگرافی GC
|
Maximum temperature: 375 °C
|
Pressure range: 0–100 psi (0–150 psi optional)
|
Accepts columns from 50-µm id to 530-µm id
|
Pressure pulse mode
|
Gas saver mode
|
Total flow ranges through the inlet: He, 20 to 1,000; H2, 26 to 1,000; N2, 20 to 200 mL/min
|
Split ratio may be adjusted electronically without affecting column flow or head pressure
|
دتکتور FIDگاز کروماتوگراف GC
|
Maximum temperature: 375 °C
|
Automatic ignition
|
Flame-out detection and reignition
|
Minimum detectable: <5 pg
carbon/s as propane using N2
carrier and a 0.29-mm id jet
|
Linear dynamic range:
Linear dynamic range: 107 (±10%)
|
Full range auto scaling
|
Data acquisition rate: up to 200 Hz
|
Flows: air, 0 to 800; hydrogen, 0 to 100; make-up (He or N2 ), 0 to 100 mL/min
|
دتکتور TCD کروماتوگرافی گازی GC
|
Maximum temperature: 375 °C
|
Minimum detectable: <400 pg propane/mL helium carrier (MDL can be affected by laboratory environment.)
|
Linear dynamic range: 105 (± 5%)
|
Flows: reference gas, 0 to 100; make-up (He, H2, N2), 1 to 12 mL/min
|
دتکتور Micro-ECD کروماتوگراف گازی GC
|
Equipped with hidden anode and high velocity flows for contamination resistance
|
400 °C maximum operating temperature
|
Makeup gas types: argon/5% methane or nitrogen
|
Radioactive source: <15 mCi63 Ni
|
MDL: <0.008 pg/s lindane
|
Dynamic range: >5 × 105 with lindane
|
Data acquisition rate: up to 50 Hz
|
دتکتور FPDگاز کروماتوگرافی GC
|
Available in single wavelength only
|
250 °C maximum operating temperature
|
MDL: <20 pg S/s, <0.9 pg P/s with dodecanethiol/tributylphosphate mixture
|
Selectivity: 105 gS/gC, 106 gP/gC
|
Dynamic range: >103 S, 104 P with dodecanethiol/tributylphosphate mixture
|
رابط ASLگاز کروماتوگراف GC
|
7683 ALS Interface standard. Provides power and communications for up to two 7683 automatic injectors, one automatic sampler tray, and one bar code reader.
|
Injector and tray install easily without the need for alignment.
|
PCOCکروماتوگرافی گازی GC
|
Maximum temperature: 375 °C.
|
Three temperature program ramps or oven track mode
|
Maximum temperature ramp rate: 350 °C/min
|
Pressure setting range: 0–100 psi
|
Total flow setting range: 0 to 100 mL/min
|
Available with LCO2 cyrogenic cooling only.
|
Lower temperature limit without LCO2 cyrogenic cooling: Oven temperature plus 3 °C with oven track on, –20 °C with oven track off
|
Lower temperature with LCO2 cyrogenic cooling: -17 °C with oven track on -20 °C with oven tracks off
|
PTVکروماتوگراف گاز GC
|
Maximum temperature: 375 °C
|
Three temperature program ramps
|
Temperature ramp rates 0.1 to 720 °C/min
|
Pressure setting range: 0–100 psi
|
Total flow setting range: 0 to 200 mL/min N2 0 to 1,000 mL/Min H2 or He
|
Available with LCO2 cyrogenic cooling only
|
Lower temperature limit without CO2: ambient plus 10 °C
|
Lower temperature with LCO2 cyrogenic cooling: –30 °C (10 °C less than oven)
|
Available with Gerstel septumless head or Merlin MicroSeal septum head
|
ابعاد کروماتوگرافی گازی GC
|
Height: 490 mm to FID cover top, and 505 mm to top of valve box
|
ولوهای قابل پشتیبانیگاز کروماتوگراف GC
|
One 6-port gas sampling valve or one 4-port liquid sampling valve can be mounted in a heated compartment. The valve can be connected to the column either directly or via the inlet. Valves may be run-time programmed. Actuating a sampling valve will START a run. Both the traditional actuator/rotary valves as well as piston/diaphragm valves are available. The rotary valves are suitable for most applications. The air-actuated diaphragm valves have a much longer life and are easier to maintain; however, they are not suitable for ammonia, primary and secondary amines, hydrazines, strong oxidizers, and alkaline solutions (pH >10). The recommended operating temperature and pressure ranges for the miniature VICI® diaphragm design are: 50 to 150 °C and 20–300 psi. The valves are actuated pneumatically and require 40 to 50 psi air pressure. The recommended operating temperature range for the rotary 6-port gas sampling valve is 50 to 200 °C and the maximum sample pressure is 400 psi. The operating temperature range for the rotary 4-port liquid sampling valves is 70 °C to 175 °C. At 5000 psi, the maximum temperature is 75 °C while at 1000 psi the maximum temperature is 175 °C. The gas sampling valve, rotary or diaphragm, is supplied with a 0.25-cc loop. Other loops (0.5, 1, and 2 cc) are available separately. Liquid sample valves have a built-in sample loop (0.5 or 1.0-µL size).
|