Main system |
PC communication |
USB2.0 High Speed |
Line voltage |
100-240VAC, 50/60Hz, 1Amp |
Power adapter |
24V @2.5A 60Watt |
Size/weight |
411X321X165mm(W*D*H) / 4.4kg |
Max output power |
15Watt |
System |
Cell cable |
1meter shielded type (Standard) |
|
Working, reference, counter, |
|
Working sense,Auxiliary V |
Control |
DSP with FPGA |
Calibration |
Automatic |
Data acquisition |
2x16bit ADCs (500kHz)for voltage, current |
ADC |
1x 16bit ADCs(250kHz) for auxiliary voltage |
DAC |
2x16bit DAC(50MHz) for bias & scan |
|
Filter |
4ea (5Hz,1kHz,500kHz, 5MHz) |
Scan rate |
0 – 200V/sec |
0-5000V/sec(fast scan) |
Max. channel No. |
16 Channels via USB connection |
Internal Data Memory |
295,000 points |
Case |
5meter water proof |
Power amplifier(CE) |
Power |
12Watt (12V@1A) |
Compliance voltage |
±12V |
Maximum current |
±1A |
Control speed selection |
4ea |
bandwidth |
2MHz |
Slew rate |
10V/usec |
(Potentiostat mode(voltage control |
- voltage control |
Control voltage range |
±10V, ±1V, ±100mV |
Voltage resolution |
16 bit per each range |
Voltage accuracy |
±1mV±0.05% of setting |
Max. scan range |
±10V vs. ref. E |
- current measurement |
Current ranges |
120ranges(auto/manual setting) |
1nA ~ 1A |
(1nA & 10nA with gain) |
Current resolution |
16 bit; |
30uA,3uA,300nA,30nA,3nA,300pA,, |
30pA,3pA, 300fA, 30fA, |
Current accuracy |
±10pA±0.1% of range(gain x1) >100nA fs. |
Galvanostat mode(current control) |
- Current control |
Control current range |
Max. ±1A |
± full scale depending on selected range |
Current resolution |
16 bit; |
30uA,3uA,300nA,30nA,3nA,300pA,, |
30pA,3pA, 300fA, 30fA, |
|
Current accuracy |
±10pA±0.1% of range(gain x1) >100nA fs. |
- Voltage measurement |
Voltage ranges |
10V,1V,100mV |
Voltage resolution |
16 bit; 0.3mV,30uV,3uV |
Voltage accuracy |
±1mV±0.05% of reading |
Electrometer |
Max input voltage |
±10V |
Input impedance |
2x1013ヘ||4.5pF |
Bandwidth |
>22MHz |
CMRR |
>114dB |
|
|
EIS(Internal FRA) for system |
Frequency range |
10uHz ~ 1MHz |
Frequency accuracy |
0.01% |
Frequency resolution |
5000/decade |
Amplitude |
0.1mV ~ 5V rms (Potentiostatic) |
0.1 ~ 70% f.s(Galvanostatic) |
Mode |
Static EIS: |
Potentiostatic, galvanostatic, pseudo-galvanostatic, OCP |
Dynamic EIS: |
Potentiodynamic, Galvanodynamic |
Fixed frequency Impedance: |
Potentiostatic, Galvanostatic, potentiodynamic, galvanodynamic |
Multisine EIS: |
Potentiostatic, Galvanostatic |
Intermittent PEIS/GEIS |
Interfaces for system |
Auxiliary port |
|
Auxiliary voltage inputs |
1 analog Input: ±10V |
|
For measurement of WE vs. CE |
Zero Resistance Ammeter |
100nA ~ 1A ranges |
Software |
Max steps per experiment |
1000 |
Shutdown safety limits |
Voltage, current,AuxVetc |
Max. sampling rate |
20kHz(50usec)-burst mode/500jHz(2usec) in fast sweep mode |
Max. sampling time |
unlimited |
Sampling condition |
Time, dv/dt, dI/dt etc. |
PC requirement |
|
Operating system |
WindowsXP SP3/7/8/10 |
PC specification(minimum) |
Pentium4, RAM 1GB |
Display |
1600x900 high color |
USB |
High speed 2.0 |
General |
Dummy cell |
One external dummy cell included |