Main system |
PC communication |
USB2.0 High Speed |
Line voltage |
100-240VAC, 50/60Hz, 1Amp |
LED |
Run,Busy,Potentiostat,Galvanostat |
Size/weight |
179 x 270 x 378.4(W*H*D) / 7.65kg |
Power consumption |
24V@6.3A 150Watt |
System |
Cell cable |
(1meter shielded type (Standard |
|
Power terminal: Working, counter, |
|
Sensing terminal: Reference, Working sense |
Control |
DSP with FPGA |
Calibration |
Automatic |
Data acquisition |
2x16bit ADCs (500kHz)for voltage, current |
ADC |
4x 16bit ADCs(250kHz) for auxiliary voltage & temperature reading |
DAC |
2x16bit DAC(50MHz) for bias & scan |
|
1X16bit DAC(1MHz) for analog output |
Filter |
4ea (5Hz,1kHz,500kHz, 5MHz) |
Scan rate |
0 – 200V/sec |
|
0-5000V/sec(fast scan) |
Max. channel No. |
32 Channels via USB connection |
Internal Data Memory |
295,000 points |
LCD Display |
DC & EIS mode automatically |
Power amplifier(CE) |
|
Power |
50Watt (10V@5A) |
Compliance voltage |
±10V |
Maximum current |
±5A |
Control speed selection |
8ea |
bandwidth |
2MHz |
Slew rate |
10V/usec |
Potentiostat mode(voltage control) |
|
- voltage control |
|
Control voltage range |
±10V, ±1V, ±100mV |
Voltage resolution |
16 bit per each range |
Voltage accuracy |
±1mV±0.05% of setting |
Max. scan range |
±10V vs. ref. E |
- current measurement |
|
Current ranges |
12 ranges(auto/manual setting) |
|
50pA ~ 5A |
|
(50pA & 500pA with gain) |
Current resolution |
16 bit; |
|
150uA,15uA,1.5uA,150nA,15nA,1.5nA, |
|
150pA,15pA, 1.5pA, 150fA,15fA,1.5fA |
Current accuracy |
±10pA±0.1% of range(gain x1)>500nA fs |
Galvanostat mode(current control) |
|
- Current control |
|
Control current range |
Max. ±5A |
|
± full scale depending on selected range |
Current resolution |
16 bit; |
|
150uA,15uA,1.5uA,150nA,15nA,1.5nA, |
|
150pA,15pA, 1.5pA, 150fA,15fA,1.5fA |
|
|
Current accuracy |
±10pA±0.1% of range(gain x1)>500nA fs |
- Voltage measurement |
|
Voltage ranges |
10V,1V,100mV |
Voltage resolution |
16 bit; 0.3mV,30uV,3uV |
Voltage accuracy |
±1mV±0.05% of reading |
Electrometer |
|
Max input voltage |
±10V |
Input impedance |
2x1013¥Ø||4.5pF |
|
|
Bandwidth |
>22MHz |
CMRR |
>114db |
EIS(Internal FRA) for system |
|
Frequency range |
10uHz ~ 1MHz |
Frequency accuracy |
0.01% |
Frequency resolution |
5000/decade |
Amplitude |
0.1mV ~ 5V rms (Potentiostatic) |
|
0.1 ~ 70% f.s(Galvanostatic) |
Mode |
Static EIS: |
|
Potentiostatic, galvanostatic, pseudo-galvanostatic, OCP |
|
Dynamic EIS: |
|
Potentiodynamic, Galvanodynamic |
|
Fixed frequency Impedance: |
|
Potentiostatic, Galvanostatic, potentiodynamic, galvanodynamic |
|
Multisine EIS: |
|
Potentiostatic, Galvanostatic |
|
Intermittent PEIS/GEIS |
Interfaces for system |
|
Auxiliary port |
|
Digital output |
3 (open collector) |
Digital input |
2 (photo coupler) |
Auxiliary voltage inputs |
3 analog Inputs: ±10V |
|
For measurement of WE vs. CE |
|
CE vs. RE or other signal |
Analog output |
1 analog output: ±10V |
|
For stirrer, MFC, RDE etc. |
Misc. port |
|
Sig generator output |
1 analog output for FRA output or |
|
waveform generation output |
Peripheral communication |
I2C to control external devices |
|
|
Temperature measurement |
1 K-type thermocouple input |
Zero Resistance Ammeter |
5nA ~ 5A ranges |
Software |
|
Max steps per experiment |
1000 |
Shutdown safety limits |
Voltage, current, temperature etc |
Max. sampling rate |
20kHz(50usec)-burst mode/500jHz(2usec) in fast sweep mode |
Max. sampling time |
unlimited |
Sampling condition |
Time, dv/dt, dI/dt, temp etc. |
PC requirement |
|
Operating system |
WindowsXP SP3/7/8/10 |
PC specification(minimum) |
Pentium4, RAM 1GB |
Display |
1600x900 high color |
USB |
High speed 2.0 |
Dummy cell |
One external dummy cell included |
Thermocouple |
1.5meter K-type (option) |
Auxiliary cable |
Option |
Misc. cable
|