| Main system |
| PC communication |
USB2.0 High Speed |
| Line voltage |
100-240VAC, 50/60Hz, 1Amp |
| LED |
Run,Busy,Potentiostat,Galvanostat |
| Size/weight |
179 x 270 x 378.4(W*H*D) / 7.65kg |
| Power consumption |
24V@6.3A 150Watt |
| System |
| Cell cable |
(1meter shielded type (Standard |
| |
Power terminal: Working, counter, |
| |
Sensing terminal: Reference, Working sense |
| Control |
DSP with FPGA |
| Calibration |
Automatic |
| Data acquisition |
2x16bit ADCs (500kHz)for voltage, current |
| ADC |
4x 16bit ADCs(250kHz) for auxiliary voltage & temperature reading |
| DAC |
2x16bit DAC(50MHz) for bias & scan |
| |
1X16bit DAC(1MHz) for analog output |
| Filter |
4ea (5Hz,1kHz,500kHz, 5MHz) |
| Scan rate |
0 – 200V/sec |
| |
0-5000V/sec(fast scan) |
| Max. channel No. |
32 Channels via USB connection |
| Internal Data Memory |
295,000 points |
| LCD Display |
DC & EIS mode automatically |
| Power amplifier(CE) |
|
| Power |
50Watt (10V@5A) |
| Compliance voltage |
±10V |
| Maximum current |
±5A |
| Control speed selection |
8ea |
| bandwidth |
2MHz |
| Slew rate |
10V/usec |
| Potentiostat mode(voltage control) |
|
| - voltage control |
|
| Control voltage range |
±10V, ±1V, ±100mV |
| Voltage resolution |
16 bit per each range |
| Voltage accuracy |
±1mV±0.05% of setting |
| Max. scan range |
±10V vs. ref. E |
| - current measurement |
|
| Current ranges |
12 ranges(auto/manual setting) |
| |
50pA ~ 5A |
| |
(50pA & 500pA with gain) |
| Current resolution |
16 bit; |
| |
150uA,15uA,1.5uA,150nA,15nA,1.5nA, |
| |
150pA,15pA, 1.5pA, 150fA,15fA,1.5fA |
| Current accuracy |
±10pA±0.1% of range(gain x1)>500nA fs |
| Galvanostat mode(current control) |
|
| - Current control |
|
| Control current range |
Max. ±5A |
| |
± full scale depending on selected range |
| Current resolution |
16 bit; |
| |
150uA,15uA,1.5uA,150nA,15nA,1.5nA, |
| |
150pA,15pA, 1.5pA, 150fA,15fA,1.5fA |
| |
|
| Current accuracy |
±10pA±0.1% of range(gain x1)>500nA fs |
| - Voltage measurement |
|
| Voltage ranges |
10V,1V,100mV |
| Voltage resolution |
16 bit; 0.3mV,30uV,3uV |
| Voltage accuracy |
±1mV±0.05% of reading |
| Electrometer |
|
| Max input voltage |
±10V |
| Input impedance |
2x1013¥Ø||4.5pF |
| |
|
| Bandwidth |
>22MHz |
| CMRR |
>114db |
| EIS(Internal FRA) for system |
|
| Frequency range |
10uHz ~ 1MHz |
| Frequency accuracy |
0.01% |
| Frequency resolution |
5000/decade |
| Amplitude |
0.1mV ~ 5V rms (Potentiostatic) |
| |
0.1 ~ 70% f.s(Galvanostatic) |
| Mode |
Static EIS: |
| |
Potentiostatic, galvanostatic, pseudo-galvanostatic, OCP |
| |
Dynamic EIS: |
| |
Potentiodynamic, Galvanodynamic |
| |
Fixed frequency Impedance: |
| |
Potentiostatic, Galvanostatic, potentiodynamic, galvanodynamic |
| |
Multisine EIS: |
| |
Potentiostatic, Galvanostatic |
| |
Intermittent PEIS/GEIS |
| Interfaces for system |
|
| Auxiliary port |
|
| Digital output |
3 (open collector) |
| Digital input |
2 (photo coupler) |
| Auxiliary voltage inputs |
3 analog Inputs: ±10V |
| |
For measurement of WE vs. CE |
| |
CE vs. RE or other signal |
| Analog output |
1 analog output: ±10V |
| |
For stirrer, MFC, RDE etc. |
| Misc. port |
|
| Sig generator output |
1 analog output for FRA output or |
| |
waveform generation output |
| Peripheral communication |
I2C to control external devices |
| |
|
| Temperature measurement |
1 K-type thermocouple input |
| Zero Resistance Ammeter |
5nA ~ 5A ranges |
| Software |
|
| Max steps per experiment |
1000 |
| Shutdown safety limits |
Voltage, current, temperature etc |
| Max. sampling rate |
20kHz(50usec)-burst mode/500jHz(2usec) in fast sweep mode |
| Max. sampling time |
unlimited |
| Sampling condition |
Time, dv/dt, dI/dt, temp etc. |
| PC requirement |
|
| Operating system |
WindowsXP SP3/7/8/10 |
| PC specification(minimum) |
Pentium4, RAM 1GB |
| Display |
1600x900 high color |
| USB |
High speed 2.0 |
| Dummy cell |
One external dummy cell included |
| Thermocouple |
1.5meter K-type (option) |
| Auxiliary cable |
Option |
Misc. cable
|