Information

 

Potentiostat|Galvanostat

  • Model : ZIVE SP2
  • Company : WonATech
  • Country : South Korea

ZIVE SP2, the outstanding potentiostat/galvanostat /FRA, is the best choice for the complete DC and impedance characterization of various electrochemical application

Introduction

ZIVE SP2, the outstanding potentiostat/galvanostat/FRA, is the best choice for the complete DC and impedance characterization of various electrochemical application.  

Versatility

ZIVE SP2 is system comes with additional three analog inputs (auxiliary voltage input) and 1 analog output, whereas 3 digital outputs and 2 digital inputs and one temperature input for Ktype thermocouple. It will help user expand the usage of instrument.

Application

ZIVE SP2 electrochemical workstation is ideal for fundamental research in electrochemistry, development and quality assurance of new sensors, corrosion/coatings, electrode material, membrane, conducting polymer, evaluation power device research such as battery materials, fuel cells, super capacitors and solar cells.

Software (Smart Manager) 

ZIVE SP2 electrochemical workstation use Smart Manager software. You can get more detailed information via above link.

ZIVE SP2 is equipped with a frequency response analyzer(FRA) for system as standard and it provides high performance impedance measurements over the frequency range 10uHz to 1MHz.  ZRA(zero resistance ammeter) function can measure max 2 A in galvanic corrosion technique. ZIVE SP2 can be easily upgraded with adding optional application software package without returning the system to the factory for upgrade.

Specification

Main system

PC communication

USB2.0 High Speed

Line voltage

100-240VAC, 50/60Hz, 1Amp

Power adapter

24V @2.5A 60Watt

Size/weight

113X305.7X196mm(W*D*H) / 3.35kg

Max output power

30Watt

System

Cell cable

1meter shielded type (Standard)

 

Working, reference, counter,

 

Working sense

Control

DSP with FPGA

Calibration

Automatic

Data acquisition

2x16bit ADCs (500kHz)for voltage, current

ADC

4x 16bit ADCs(250kHz) for auxiliary voltage & temperature reading

DAC

2x16bit DAC(50MHz) for bias & scan

1X16bit DAC(1MHz) for analog output

Filter

4ea (5Hz,1kHz,500kHz, 5MHz)

Scan rate

0 – 200V/sec

0-5000V/sec(fast scan)

Max. channel No.

32 Channels via USB connection

Internal Data Memory

295,000 points

LCD Display

DC & EIS mode automatically

Power amplifier(CE)

Power

24Watt (12V@2A)

Compliance voltage

±12V

Maximum current

±2A

Control speed selection

8ea

bandwidth

5MHz

Slew rate

10V/usec

Potentiostat mode(voltage control)

- voltage control

Control voltage range

±10V, ±1V, ±100mV

Voltage resolution

16 bit per each range

Voltage accuracy

±1mV±0.05% of setting

Max. scan range

±10V vs. ref. E

- current measurement

Current ranges

12 ranges(auto/manual setting)

20pA ~ 2A

(20pA & 200pA with gain)

Current resolution

16 bit;

60uA,6uA,600nA,60nA,6nA,600pA,

60pA,6pA, 600fA, 60fA,6fA,0.6fA

Current accuracy

±10pA±0.1% of range(gain x1) >200nA fs

(Galvanostat mode(current control

- Current control

Control current range

Max. ±2A

± full scale depending on selected range

Current resolution

16 bit;

60uA,6uA,600nA,60nA,6nA,600pA,

60pA,6pA, 600fA, 60fA,6fA,0.6fA

¡¡

Current accuracy

±10pA±0.1% of range(gain x1) >200nA fs

- Voltage measurement

Voltage ranges

10V,1V,100mV

Voltage resolution

16 bit; 0.3mV,30uV,3uV

Voltage accuracy

±1mV±0.05% of reading

Electrometer

Max input voltage

±10V

Input impedance

2x1013¥Ø||4.5pF

Bandwidth

>22MHz

CMRR

>114db

EIS(Internal FRA) for system

Frequency range

10uHz ~ 2MHz

Frequency accuracy

0.01%

Frequency resolution

5000/decade

Amplitude

0.1mV ~ 5V rms (Potentiostatic)

0.1 ~ 70% f.s(Galvanostatic)

Mode

Static EIS

Potentiostatic, galvanostatic, pseudo-galvanostatic, OCP

Dynamic EIS:

Potentiodynamic, Galvanodynamic

Fixed frequency Impedance:

Potentiostatic, Galvanostatic, potentiodynamic, galvanodynamic

Multisine EIS

Potentiostatic, Galvanostatic

Intermittent PEIS/GEIS¡¡

Interfaces  for system

Auxiliary port

 

 Digital output

3 (open collector)

 Digital input

2 (photo coupler)

 Auxiliary voltage inputs

3 analog Inputs: ±10V

 

For measurement of WE vs. CE

 

CE vs. RE or other signal

 Analog output

1 analog output: ±10V

 

For stirrer, MFC, RDE etc.

Misc.  port

 

Sig generator output

1 analog output for FRA output or

 

 waveform generation output

Peripheral communication

I2C to control external devices

 

 

Temperature measurement

1 K-type thermocouple input

Zero Resistance Ammeter

2nA ~ 2A ranges

Software

Max steps per experiment

1000

Shutdown safety limits

Voltage, current, temperature etc

Max. sampling rate

20kHz(50usec)-burst mode/500jHz(2usec) in fast sweep mode

Max. sampling time

unlimited

Sampling condition

Time, dv/dt, dI/dt, temp etc.

PC requirement

¡¡

Operating system

WindowsXP SP3/7/8/10

PC specification(minimum)

Pentium4, RAM 1GB

Display

1600x900 high color

USB

High speed 2.0

General

Dummy cell

One external dummy cell included

Thermocouple

1.5meter K-type (option)

Auxiliary cable

Option

Misc. cable

 

Product Related

Post a comment
Review title:
your name :
email :
10 / 10
از 1 کاربر