Information

 

Potentiostat|Galvanostat 

  • Model : ZIVE BP2C
  • Company : WonATech
  • Country : South Korea

ZIVE BP2C ,  the dual channel potentiostat / galvanostat /ZRA/FRA, is to support dual working electrodes cell with one reference and one counter electrode configuration ( bi-potentiostat).

Introduction

The ZIVE BP2C, a dual channel potentiostat/galvanostat/FRA, is to support dual-working-electrode cell with one reference and one counter electrode configuration(bi-potentiostat) for sample characterization. Each channel can conduct DC and impedance test simultaneously and/or independently. The ZIVE BP2C can be setup to run 2-electrode, 3-electrode, or 4-electrode measurements with a simple setup change.

The ZIVE BP2C’s each channel is equipped with a frequency response analyzer(FRA) as standard and it provides high performance impedance measurements over the frequency range 10uHz to 1MHz. ZRA(zero resistance ammeter) function can measure max. 2 Amp in galvanic corrosion technique. The system is supplied with four(4) advanced software packages, which are catagorized by application fields. With this advanced software packages, user can widen ZIVE BP2C’s flexibility.

features

Features System

  • Versatile high quality dual channel potentiostat/galvanostat/impedance analyzer
  • Bi-potentiostat
  1. two fully independent channels
  2. dual working electrodes with one reference and one counter electrode
  3. configuration available
  • FRA function to control an external electronic load or 3rd party potentiostat/galvanostat is available as standard
  • 14 EIS techniques capability including multisine
  • Current interrupt IR measurement IR compensation(dynamic, positive feedback)
  • Bipolar pulse capability
  • Voltage pulse or current pulse charge/discharge test(GSM,CDMA etc.), sine wave function for ripple simulation in battery test package and pulse plating available
  • High speed data sampling time per channel
  1.  50usec/sample in burst mode
  2.  1msec/sample in normal mode
  3. 2usec/sample in fast sweep mode
  • Fast sweep mode(5000V/sec with 10mV data sampling)
  • 3 measurement/control voltage ranges & 12 measurement/control current ranges

Application

Application:

The ZIVE BP2C, a dual channel electrochemical workstation is ideal for bio sensor research, FET sensor, permeation test etc. And it can be also used for fundamental research in electrochemistry, development and quality assurance of new sensors, corrosion/ coatings, electrode material, membrane, conducting polymer, evaluation power device research such as battery materials, fuel cells, super capacitors and solar cells.

Specification

Specificaton:

Main system

 

PC communication

USB2.0 High Speed

 

Line voltage

100-240VAC, 50/60Hz, 1Amp

 

Max. channel number

2 independent channels per unit

 

Size/weight

209X270X378mm(WxHxD) / 9.25kg

 

Max output power

30Watt

 

System

 

Cell cable

1meter shielded type (Standard)

 

 

Working, reference, counter,

 

 

Working sense

 

Control

DSP with FPGA

 

Calibration

Automatic

 

Data acquisition

2x16bit ADCs (500kHz)for voltage, current

 

ADC

4x 16bit ADCs(250kHz) for auxiliary voltage & temperature reading

 

DAC

2x16bit DAC(50MHz) for bias & scan

 

1X16bit DAC(1MHz) for analog output

 

Filter

4ea (5Hz,1kHz,500kHz, 5MHz)

 

Scan rate

0 – 200V/sec
0-5000V/sec(fast scan)

 

Max. channel No.

32 Channels via USB connection

 

Internal Data Memory

295,000 points

 

LCD Display

DC & EIS mode automatically

 

Power amplifier(CE)

 

Power

24Watt (12V@2A)

 

Compliance voltage

±12V

 

Maximum current

±2A

 

Control speed selection

8ea

 

bandwidth

5MHz

 

Slew rate

10V/usec

 

Potentiostat mode(voltage control)

 

- voltage control

 

Control voltage range

±10V, ±1V, ±100mV

 

Voltage resolution

16 bit per each range

 

Voltage accuracy

±1mV±0.05% of setting

 

Max. scan range

±10V vs. ref. E

 

- current measurement

 

Current ranges

12 ranges(auto/manual setting)

 

20pA ~ 2A

 

(20pA & 200pA with gain)

 

Current resolution

16 bit;

 

60uA,6uA,600nA,60nA,6nA,600pA,

 

60pA,6pA, 600fA, 60fA,6fA,0.6fA

 

Current accuracy

±10pA±0.1% of range(gain x1) >200nA fs

 

Galvanostat mode(current control)

 

- Current control

 

Control current range

Max. ±2A

 

± full scale depending on selected range

 

Current resolution

16 bit;

 

60uA,6uA,600nA,60nA,6nA,600pA,

 

60pA,6pA, 600fA, 60fA,6fA,0.6fA

 

 

 

Current accuracy

±10pA±0.1% of range(gain x1) >200nA fs

 

- Voltage measurement

 

Voltage ranges

10V,1V,100mV

 

Voltage resolution

16 bit; 0.3mV,30uV,3uV

 

Voltage accuracy

±1mV±0.05% of reading

 

Electrometer

 

Max input voltage

±10V

 

Input impedance

2x1013¥Ø||4.5pF

 
 

Bandwidth

>22MHz

 

CMRR

>114db

 

EIS(Internal FRA) for system

 

Frequency range

10uHz ~ 2MHz

 

Frequency accuracy

0.01%

 

Frequency resolution

5000/decade

 

Amplitude

0.1mV ~ 5V rms (Potentiostatic)

 

0.1 ~ 70% f.s(Galvanostatic)

 

Mode

Static EIS:

 

Potentiostatic, galvanostatic, pseudo-galvanostatic, OCP

 

Dynamic EIS:

 

Potentiodynamic, Galvanodynamic

 

Fixed frequency Impedance:

 

Potentiostatic, Galvanostatic, potentiodynamic, galvanodynamic

 

Multisine EIS:

 

Potentiostatic, Galvanostatic

 

Intermittent PEIS/GEIS¡¡

 

Interfaces  for system

 

Auxiliary port

 

 

Digital output

3 (open collector)

 

Digital input

2 (photo coupler)

 

Auxiliary voltage inputs

3 analog Inputs: ±10V

 

 

For measurement of WE vs. CE

 

 

CE vs. RE or other signal

 

Analog output

1 analog output: ±10V

 

 

For stirrer, MFC, RDE etc.

 

Misc.  port

 

 

Sig generator output

1 analog output for FRA output or

 

 

waveform generation output

 

Peripheral communication

I2C to control external devices

 

   

 

Temperature measurement

1 K-type thermocouple input

 

Zero Resistance Ammeter

2nA ~ 2A ranges

 

Software

 

Max steps per experiment

1000

 

Shutdown safety limits

Voltage, current, temperature etc

 

Max. sampling rate

20kHz(50usec)-burst mode/500jHz(2usec) in fast sweep mode

 

Max. sampling time

unlimited

 

Sampling condition

Time, dv/dt, dI/dt, temp etc.

 

PC requirement

 

 

Operating system

WindowsXP SP3/7/8/10

 

PC specification(minimum)

Pentium4, RAM 1GB

 

Display

1600x900 high color

 

USB

High speed 2.0

 

General

 

Dummy cell

One external dummy cell included

 

Thermocouple

1.5meter K-type (option)

 

Auxiliary cable

Option

 

Misc. cable

Option

 

 

Product Related