Information

 

Gel permeation Chromatography ( SEC ) system equipped

  • Model : HLC-8320GPC
  • Company : Tosoh
  • Country : Japan

HLC-8320GPC is a dedicated size exclusion chromatography (SEC) system equipped with the functions necessary for SEC analysis. And the GPC Workstation EcoSEC-WS gives full control of the main unit, enabling molecular mass analysis with excellent operability and high precision.

This system complies with various regulations (CE, UL, FDA 21 CFR Part 11).

Introduction

Features

  • High Reliability

When the system is used with semi-micro SEC columns, featuring calibration curves with superior

linearity*, and the standard polystyrene kit  PStQuick, which ensures minimization of  preparation errors, highly reliable data with minimized variation between systems may be obtained.

(*Compared with Tosoh's conventional products)

  • High Stability

The column oven has a double structure and a double temperature-control mechanism.

A double-path, double-flow refractive index detector (RI detector) is used for temperature control of the main units, including tubing. A stable baseline is obtained by suppressing fluctuation of the refractive index in the solvent.

  • High Reproducibility

The pumping unit features a newly designed structure and control system. Temperature control of the

entire pumping unit gives high reproducibility which is not affected by varying solvent conditions or environmental temperature fluctuations.

  • Fast, Efficient, Low Solvent Consumption

Combined use with TSKgel semi-micro SEC columns (SuperMultiporeHZ, SuperHZ, SuperH, and

SuperAW) allows fast, efficient separation and SEC analysis and reduces solvent consumption.

  • Easy Operation

Using  the  GPC  Workstation  EcoSEC-WS,  the  system  can  easily  be  operated  through  warm-up,

analysis, and shut-down. The main unit may also be independently controlled using the operator panel. The automatic purge function eliminates solvent replacement

and other time-consuming manual operations.

Global Standard

Compliant with China RoHS.

UL61010-1 and CE mark certification.

Also compliant with FDA 21 CFR PART 11*.

(*Authentication using user ID and password, logout when finished, software validation, etc.)

Options

     UV-8320

No extra installation space is necessary for this built-in option. As it can be used in the pump oven, the detector is highly stable.

Specification

Specifications

Description

Method

Dual beam, single flow cell

Wavelength

195 – 350 nm

Flow Cell

Volume

2 µL

 

Maximum pressure

3 MPa

 

Noise

2.5 × 10-5 ABU

 

 

(254 nm, cell containing air, response: 1.0 s)

 

Drift

3.0 ×10-4 ABU/h

 

 

(254 nm, cell containing air, response: 1.0 s)

Safety Features

Leak sensor, lamp lighting monitor

 

Pan for liquid leakage in the installation area

Weight

7 kg

Product Related