Dimensions: 28 × 31 × 16 cm. Accommodates up to two 105 m × 0.530 mm id capillary columns or two 10-ft glass packed columns (9 in coil diameter, 1/4 in od), or two 20-ft stainless steel packed columns (1/8 in od).
|
Operating temperature range suitable for all columns and chromatographic separations. Ambient temperature +4 °C to 450 °C.
With LN2 cryo cooling: –80 to 450 °C.
With CO2 cryo cooling: –40 to 450°C
|
Temperature set point resolution: 1 °C.
|
Supports 6 oven ramps with 7 plateaus. Negative ramps are allowed.
|
Maximum achievable temperature ramp rate: 120 °C/min (120 V units are limited to 75 °C/min, see Table 1).
|
Maximum run time: 999.99 min (16.7 h).
|
Oven cool down (22 °C ambient) 300 °C to 50 °C in 4.5 min).
|
Ambient temperature rejection:
|
Support for up to 8 valves. - Valves 1-4, 12 V DC 13 watt in a heated valve box - Valves 5-6, 24 V DC 100 mA unheated, for low power valve applications - Valves 7–8, externally powered as a remote event from separate contact closure
|
Independent heated zones, not including oven: six (two inlets, two detectors, and two auxiliary).
|
Maximum operating temperatures for auxiliary zones: 400 °C.
|
کنترل الکترونیکی پنوماتیک ها (EPC) کروماتوگراف گازی GC
|
Atmospheric pressure compensation sensor to compensate for altitude and ambient temperature variation is standard.
|
Pressure may be adjusted by increments of 0.01 psi.
|
Pressure/Flow ramps: Three maximum.
|
Carrier and makeup gas settings selectable for He, H2, N2, and argon/methane
|
Flow or pressure setpoints for each inlet or detector parameter with both Agilent 6890N and Agilent ChemStations.
|
Automated carrier has constant flow, when capillary column dimensions are entered into the 6890.
|
Split/Splitless, PTV inlets have flow sensors for the control of split ratio.
|
Inlet modules Pressure sensors: Accuracy:
Pressure sensors: Accuracy:
<±2% full scale, Repeatability:
<±0.05 psi, Temperature coefficient: <±0.01 psi/°C, Drift:
<±0.1 psi/6 months.
Flow sensors: Accuracy: <±5%
depending on carrier gas,
Repeatability: <±0.35% of setpoint, Temperature coefficient:
<±0.20 mL/min normalized temperature and pressure (NTP)*
per °C for He or H2;
<±0.05 mL/min NTP per °C for
N2 or Ar/CH4.
|
Detector modules
Accuracy: <±3 mL/min NTP or
7% of setpoint, Repeatability:
<±0.35% of setpoint, Temperature
coefficient: <±0.20 mL/min NTP
per °C
|
ورودی های کروماتوگراف گاز GC
|
Maximum of two inlets installed.
|
Inlets available:
- Packed purged injection port (PPIP)
- Split/Splitless capillary inlet (S/SL)
- Temperature-programmable cool on-column (PCOC)
- Programmable temperature vaporizer (PTV)
- Volatiles inlet (VI)
|
Split/Splitless گاز کروماتوگرافی GC
|
Suitable for all capillary columns.
|
Split ratios up to 7500:1 to avoid column overload.
|
Splitless mode for trace analysis. Pressure-pulsed splitless is easily accessible for best performance.
|
Maximum temperature: 400 °C.
|
EPC available in two pressure ranges: 0–100 psig (0 to 680 kPa) for best control for columns 0.200 mm diameter; 0–150 psig for columns
|
Gas saver mode to reduce gas consumption without compromising performance.
|
Optimized septum purge flow to eliminate “ghost” peaks.
|
Total flow setting range: 0 to 200 mL/min N2 0 to 1,000 mL/min H2 or He
|
Flip top accessory is available for quick, easy, injector liner changes.
|
Available with manual pneumatics.
|
دتکتور FID گاز کروماتوگراف GC
|
Minimum detectable level (for tridecane):
|
Linear dynamic range: >107 (±10%). Digital data makes entire range available without range changes.
|
Data rates up to 200 Hz accommodate peaks as narrow as 25 ms at half height.
|
Standard EPC for three gases:
- Air: 0 to 800 mL/min
- H2: 0 to 100 mL/min
- Makeup gas (N2 or He): 0 to 100 mL/min
|
Available in two versions: capillary column optimized or adaptable for either packed or capillary columns.
|
Flame out detection and automatic reignition.
|
Grounded jet.
|
450 °C maximum operating temperature.
|
دتکتور TCD کروماتوگرافی گازی GC
|
Minimum detectable level: 400 pg propane/mL with He carrier. (This value may be affected by laboratory environment.)
|
Linear dynamic range: >105 ±5%
|
Unique fluidic switching design provides rapid stabilization from turn-on, low-drift performance.
|
Signal polarity can be run programmed for components having higher thermal conductivity than the carrier gas.
|
Maximum temperature: 400 °C.
|
Standard EPC for 2 gases (He, H2, Ar, or N2 – matched to carrier gas type).
|
Makeup gas: 0 to 12 mL/min
|
Reference gas: 0 to 100 mL/min
|
دتکتور Micro-ECD کروماتوگراف گازی GC
|
Minimum detectable level:
|
Proprietary signal linearization Linear dynamic range: >5 × 104 with lindane.
|
Data acquisition rate: up to 50 Hz.
|
Uses ß emission of <15 mCi 63Ni
as the electron source.
|
Unique micro-cell design minimizes contamination and
optimizes sensitivity.
|
400 °C maximum operating temperature.
|
Standard EPC makeup gas types: argon/5% methane or nitrogen; 0 to 150 mL/min.
|
Dynamic range: >5 × 105 with
lindane.
|
دتکتور NPD کروماتوگراف گاز GC
|
MDL: <0.4 pg N/s, <0.2 pg P/s with azobenzene/malathion/ octadecane mixture.
|
Dynamic range: >105 N, >105 P with azobenzene/malathion/ octadecane mixture.
|
Selectivity: 25,000 to 1 gN/gC, 75,000 to 1 gP/gC with azobenzene/malathion/octadecane mixture.
|
Data acquisition rate: up to 200 Hz.
|
Standard EPC for three gases:
- Air: 0 to 200 mL/min
- H2: 0 to 30 mL/min
- Makeup gas: 0 to 100 mL/min
|
Available for packed/capillary columns or optimized for capillary columns.
|
400 °C maximum operating temperature.
|
دتکتور FPD گاز کروماتوگرافی GC
|
MDL: <60 fg P/s, <3.6 pg S/s with methylparathion.
|
Dynamic range: >103 S, 104 P with methylparathion
|
Selectivity: 106 gS/gC, 106 gP/gC.
|
Data acquisition rate: up to 200 Hz
|
Standard EPC for three gases:
- Air: 0 to 130 mL/min
- H2: 0 to 250 mL/min
- Makeup gas: 0 to 130 mL/min
|
Available in single- or dual wavelength versions.
|
250 °C maximum operating temperature.
|
رابط ASL گاز کروماتوگراف GC
|
7683 ALS Interface standard. Provides power and communications for up to two 7683 automatic injectors, one automatic sampler tray, and one bar code reader.
|
Injector and tray install easily without the need for alignment.
|
PCOC کروماتوگرافی گازی GC
|
Direct injection onto cool capillary column assures quantitative sample transfer with no thermal degradation.
|
Automatic liquid injection supported directly onto columns ≥0.250 mm id.
|
Maximum temperature: 450 °C. Temperature programming in 3 ramps or tracking oven. Subambient control to –40 °C is optional.
|
Electronic pressure control range: 0.00 to 100.00 psig.
|
Optimized septum purge flow.
|
Optional Solvent Vapor Exit for large volume injections.
Electronically controlled, inert, three-way valve allows solvent venting.
Includes software for method optimization.
Pre-assembled retention gaps/vent line/analytical column for easy installation.
|
PPIP کروماتوگراف گازی GC
|
Direct injection onto packed and wide bore capillary columns.
|
Electronic flow/pressure control: 0.0 to 100 psig pressure range, 0.0 to 100.0 mL/min. Ranges are chosen to provide optimum performance over normal packed column set point ranges.
|
400 °C maximum operating temperature.
|
Adapters included for 1/4-inch and 1/8-inch packed columns, and 0.530-mm capillary columns.
|
Manual pneumatics available.
|
PTV کروماتوگراف گاز GC
|
Most versatile inlet for difficult samples supporting cool injections and large volume injections in split and splitless modes.
|
Temperature control: either LN2 (to –160 °C) or LCO2 (to –65 °C) cooling. Temperature programming of up to 3 ramps at up to 720 °C/min. Maximum temperature: 450 °C.
|
EPC. Pressure range 0.00 to 100 psig.
|
Split ratio up to 1000:1.
|
Choice of Gerstel septumless head or Merlin Microseal® septum head.
|
450 °C maximum operating temperature.
|
Total flow setting range:
- 0 to 200 mL/min N2
- 0 to 1,000 mL/min H2 or He
|
VI گاز کروماتوگرافی GC
|
Very low volume (32 µL) interface suitable for gas or prevaporized samples. Recommended for use with headspace, purge and trap, or thermal desorption samplers.
|
Three modes for optimized sample introduction: split (up to 100:1 split ratio), splitless, and direct.
|
Optimized EPC (H2 or He carrier, 0.00 to 100 psig pressure control, 0.0 to 100 mL/min flow control).
|
Silcosteel® treated flow path provides inert surface for minimum component adsorption.
|
Maximum temperature: 400 °C
|
ابعاد کروماتوگرافی گازی GC
|
28 × 31 × 16 cm
Accommodates up to two 105 m × 0.530 mm id capillary columns or two 10-ft glass packed columns (9 in coil diameter, 1/4 in od), or two 20-ft stainless steel packed columns (1/8 in od)
|
محدوده دمای عملیاتی کروماتوگراف گازی GC
|
Ambient temperature +4 °C to 450 °C
|
رمپ های دمایی کروماتوگراف گاز GC
|
Supports 6 oven ramps with 7 plateaus. Negative ramps are allowed
|
حداکثر نرخ رمپ دمایی قابل دست یابی گاز کروماتوگرافی GC
|
120 °C/min (120 V units are limited to 75 °C/min
|
ولوهای قابل پشتیبانی گاز کروماتوگراف GC
|
Support for up to 8 valves. – Valves 1-4, 12 V DC 13 watt in a heated valve box – Valves 5-6, 24 V DC 100 mA unheated, for low power valve applications – Valves 7–8, externally powered as a remote event from separate contact closure
|